To Set Thermocouple Type
1. Press MENU until the meter displays:
INPUT
2. Press /MAX until the meter flashes:
TC

3.
Press MENU. The meter displays:
STORED
TC

4.
Press /MAX until the meter displays the thermocouple type: (on right)
Your choices are J, K T, E, N, DNJ, R, S, and B.

JTC
5. Press MENU. The meter displays:
STORED
RDG.CNF

6.
Press RESET twice. The meter displays: and then shows the current temperature.
RESET2

To Set Temperature Units

1. Press MENU until the meter displays:
RDG.CNF
2. Press /MIN until the meter displays:
RDG.S=0
3. Press /MAX to change the value. Your choices are: 0 = ░C, 1 = ░F, 2 = ░K

4.
Press MENU to enter the value. The meter displays:
STORED
RDG.SC
RD.SC.OF

5.
Press RESET twice. The meter displays: and then shows the current temperature.
RESET2

To Set the Decimal Point

1. Press MENU until the meter displays:
2. Press /MAX to move the decimal point to the desired location.
DEC.PT
FFFF.FF
3. Press MENU. The meter displays:
STORED
CNY BY

4.
Press RESET twice. The meter displays: and then shows the current temperature.
The INFT unit is now up and running.
RESET2

 Home Home  Table of Contents TOC  Back Back